Customization: | Available |
---|---|
Capacity: | Two Person |
After-sales Service Provided: | Can Discuss |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Clean bench installation requirements:
1. This equipment should be installed in all levels of purification plant or sealed air conditioning plant.
2. This equipment should be installed away from high-speed dust sources and sources.
3. After installation, clean the inside of the equipment and its surroundings with a vacuum cleaner or a non-fiber-generating tool.
4. After the installation is cleaned, the QDF-2 type hot-ball anemometer can be used to measure the average wind speed in the clean work area. When the average wind speed is less than 0.32m/s, increase the regulator voltage and increase the fan input voltage. On the contrary, when the average wind speed is higher than 0.48m/s, turn down the regulator voltage and reduce the fan input voltage until the average value. The wind speed is kept within the ideal range of 0.4m/s ± 20% before it can be used.
Clean bench features:
1).The single-sided vertical flow closes the countertop, effectively preventing the intrusion of outside air and the odor of the operating area from harming the human body;
2).LCD intelligent controller, the control is more accurate;
3).Can set sterilization delay time, sterilization pre-opening time, ambient temperature display, ambient temperature correction, date display and adjustment, sterilization lighting can be interlocked;
3).Can query the accumulated working time and remaining effective working time;
4).Five levels of wind speed are adjustable;
5).Has the function of replacing the high efficiency filter;
6).Sliding door: double-layer suspended glass door, open up and down, rubber cushion is embedded in the bottom of the sliding door, can be stopped at any position, convenient and flexible operation;
7).Provide anti-splash built-in power socket, can directly open the front cover, easy to replace the filter element, electronic components, etc., maintenance is more convenient;
Laminar flow clean bench standard Model Technology Parameter
Product Name |
Single horizontal clean bench |
Double horizontal clean bench |
Double vertical clean bench |
Model |
YN-VC-001 |
YN-VC-002 |
YN-VC-003 |
(W×D×H)mm External Dimension |
1000*700*1800 |
1200*700*1800 |
1500*700*1800 |
(W×D×H)mm Internal Work Zone |
900*650*600 |
1100*650*600 |
1400*650*600 |
Clean Class Level |
100Class 0.3µm(USA Federal Standard) |
||
Rated Air Flow |
1200m3/h |
1800m3/h |
2300m3/h |
Air Velocity |
0.3-0.6m/s |
0.3-0.6m/s |
0.3-0.6m/s |
HEPA Efficiency |
99.99% 0.3µm H13-H14 |
||
LUX |
≥300lux |
≥300lux |
≥300lux |
Power |
124W |
127W |
200W |
Noise |
≤65dB |
≤65dB |
≤65dB |
Power Supply |
220V/50Hz |
220V/50Hz |
220V/50Hz |
LED light |
9W*1 |
13W*1 |
18W*1 |
Material |
Cabinet: epoxy powder coated steel or stainless steel, table-board: stainless steel |
Customized design is welcome.